Technology Specification
Main body:High-strength engineering plastic rack, light,
beautiful and corrosion resistant
Screen:High-definition 5.7” LCD screen display
Part of Anesthesia Machine
Vaporizer:
TwoVaporizers from Halothane, Enflurane, Isoflurane, and Sevoflurane
Automatic compensation for pressure, temperature and flow rate.
Gas Source:O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
Flow meter
High-definition four-tube flowmeter
O2 : 0.1~1.0L/min 1.1~10L/min
N20: 0.1~1.0L/min 1.1~10L/min
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off
O2 flush:25-75L/min
Low oxygen pressure alarm
There will be sound alarm when the oxygen pressure < 0.2MPa
Respiratory circuit
working mode: all-close, semi-close, semi-open
Electronic PEEP:Yes
APL:0.5~7 kPa
Respiratory Bellows
bellows for adults, tidal volume range:0~1500 ml
bellows for children, tidal volume range:0~300 ml
anyone can be chosen for application as required by the customer
Part of Ventilator
Ventilation mode:IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
Ventilation parameters
Tidal volume 50~1500 ml
Rate 2~99 bpm
SIMV rate 2~20 bpm
I:E 2:1~1:8
Inspiratory trigger pressure -1.0~2.0 kPa
PEEP 0~2.0 kPa
Pressure range 1.0~6.0 kPa
SIGH 1.5 times the inspiratory time
Parameters for ventilation monitoring
TV,ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform,flow rate - time waveform, PEEP, inspiratory trigger pressure
FiO2:21%~100%
Security Alarm System
1)Oxygen Concentration alarm
upper limit setting range 21%~100%
low limit setting range 10%~80%
2)Airway pressure alarm
upper limit setting range 0.3~6.0 kPa
low limit setting range 0.2~5.0 kPa
3)Per-minute ventilation volume alarm
upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
4)Sustained high-pressure alarm
it will give alarm when stress have consistently been higher than 2.5 kPa
5)Suffocation alarm
it will give sound and light alarm if there is no tidal volume input for 15 seconds
6)Power alarm
AC 220V±10% 50Hz±1Hz
(UPS,Storage battery)
|