Technology Specification
Screen:5 inch high-resolution LCD with touch screen
Specifications
Following is the specifications, which has complete configuration of ECG,TEMP, NIBP and SpO2.
1. General Specifications:
Size: 209mm×104mm×47.5mm
Net Weight: 655g
Display Specification
Measurement Display: color TFT- LCD/touch screen
Power Supply Indicator: Dual-color LED (red/green)
Charging indicator: A green LED
2. Power Supply
AC Power Adaptor
Input: 100~240VAC, 50/60Hz,0.7A
Output: DC 5V, 1.5A
Battery
Li-ion rechargeable battery, 3.7V,/7200mAH
Operating Time: 15 hours with full battery capacity.
(Condition: 25℃, NIBP working period is 15 minutes)
Recharge Time: 10 hours.
3. Environment
Temperature
Working:0℃~45℃
Transport and Storage: -25℃~85℃
Humidity
Working:30%~85 %
Transport and Storage: 30%~95 % (no condensation)
4. Monitoring Parameter
ECG
Input:3- lead ECG cable, standard AAMI cable connector
Lead Selection:I,II,III
Gain Selection:×1,×2,×4,×8
Sweep Speed: 12.5, 25, and 50 mm/s
Frequency Response: None filter: 0.05~100Hz (-3dB), filter: 0.5~40Hz (-3dB)
Calibration Signal: 1mV square wave
Protection: Against electrosurgical interference and defibrillation
Heart Rate
Measuring Range: 20 bpm~300 bpm
Renew Time: 1 time per beat
Accuracy: ±1 bpm
Alarm Mode: Audible and visual alarm
SpO2
SpO2%
Measuring Range:0%~100%
Resolution:1 %
Accuracy:±2 %(90%~99%), ±3 %(70%~89%),unspecified (0%~69%)
Pulse Rate
Measuring and Alarm Range:20~300bpm
Resolution: 1bpm
Accuracy: ±1bpm
NIBP
Method: Oscillometric
Mode: Manual, Auto, STAT
Measuring Interval in AUTO Mode: 1, 2, 3, 5, 10, 15, 20, 30, 45, 60 (min)
Measuring Period in STAT Mode: 5 min
Cuff Pressure Range:0 mmHg~300mmHg
Pulse Rate Range:30 bpm~250bpm
Measurement Rage: Adult 30 mmHg~255mmHg, Pediatric 30 mmHg~160mmHg
Over-pressure Protection: Adult 300mmHg, Pediatric 220mmHg
Pressure Resolution : 1mmHg
The SYSTOLIC and DIASTOLIC values measured using this device shall be
equal to those measured by trained medical personnel using cuff and stethoscope.
The MEAN value measured using this device is equal to that measured using blood pressure measurement device.
Temperature
Input: Body surface thermal-sensitive resistor temperature sensor
Measuring Range: 0℃~50℃
Accuracy: ±0.1℃
5. Storage and Review
Maximum Data Quantity: 3888 groups
Maximum Trend Time: Continuous 129.6 hours
Packing list
- Color palm patient monitor main unit 1 pc
- 3-lead ECG cable ( include surface TEMP probe) 1pc
- Adult finger SpO2 sensor 1 pc
- Adult NIBP cuff 1 pc
- NIBP extension tube 1 pc
- PC software CD 1 pc
- BlueTooth Radio 1pc
- USB cable 1 pc
- AC power adaptor 1 pc
- The operator’s manual 1 pc
Standard Configuration
ECG, HR/PR,NIBP, SpO2, RESP, TEMP
Optional Configuration
Selectable Accessory
- Pediatric finger Spo2 sensor
- Neonatal wrap SpO2 sensor
- Bluetooth printer
|